Purpose:
To evaluate and compare the relative density values of a new endodontic core material (Resilon) and conventional gutta-percha.
Methods:
15 Resilon and gutta-percha discs for testing radiopacity were prepared by first coiling three size 35 cones. The coiled material was evenly compressed into the well of a brass mold 10 mm in diameter and a depth of 1 mm. All radiographs were taken on No. 4 ultra-speed dental film. The radiopacity of Resilon, gutta-percha, and dentin slices were then measured with a digital transmission densitometer.
Results:
The mean radiopacities of Resilon and gutta-percha coiled discs were 4.06 and 7.21 mm of aluminum equivalence respectively, which exceeded the minimal requirement of 3 mm. Both tested materials had lower radiopacity values than dentin. There were highly significant differences in radiopacity values among the tested specimens (P < 0.01). Additionally, the mean radiopacity values in root canals were 5.50 and 8.52 mm Al for Resilon and gutta-percha, respectively (P < 0.01).
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Porosity in Cement Paste
The balance of water to cement in the mix is critical—it...
Porosity and Absorption of Aggregate
When all pores in an aggregate are filled with water, the aggregate is considered saturated and surface-dry. If left in dry air, water will evaporate until the aggregate...
Tooth Anatomy
The Crown, Neck, and Root
The visible part of the tooth is referred to as the crown. It's covered by enamel, the hardest substance in the human body. The crown is uniquely shaped for each type of tooth, allowing for different functions such as cutting, tearing, or grinding food.
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...