Related Experiment Video
Updated: Jun 21, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Design and development of a portable gamma radiation monitor
1Health Physics Division, Directorate of Systems and Services, Pakistan Institute of Nuclear Science and Technology (PINSTECH), Nilore, Islamabad 45650, Pakistan. mmunirali@yahoo.com
Abstract:
A portable gamma radiation monitor has been designed and developed. The monitor can be used effectively in the dose range from 0.07 to 500 mGy/h due to gamma rays of energy greater than 65 keV. The monitor overestimated radiation doses and the uncertainty in the measured dose rate has been found to be < or = 30%. The response of the monitor can be considered isotropic within an acceptable error of +/-30%. Provision has also been added to use the monitor as an installed radiation monitor. In installed mode, it can be operated from a remote location up to 1 km and the timing history can be stored on a personal computer.

![Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F55537.jpg&w=3840&q=50)