Intra-examiner repeatability and agreement in accommodative response measurements
Summary
Nott retinoscopy and binocular crossed cylinder (BCC) tests show the best repeatability for measuring accommodative response (AR). While Nott and MEM retinoscopy agree well, MEM has low repeatability, and other methods are not interchangeable for clinical use.
Area of Science:
- Ophthalmology
- Optometry
- Vision Science
Background:
- Clinical measurement of accommodative response (AR) is crucial for understanding visual function.
- Determining the repeatability of AR measurements is essential for reliable clinical assessment.
- Several methods exist for measuring AR, including retinoscopy and autorefractometry.
Purpose of the Study:
- To assess the intraexaminer repeatability of four clinical methods for measuring AR: Nott retinoscopy, monocular estimate method (MEM) retinoscopy, binocular crossed cylinder (BCC) test, and near autorefractometry.
- To evaluate the agreement between AR measurements obtained using these different clinical techniques.
Main Methods:
- Sixty-one visually normal subjects (mean age 19.7 years) underwent repeated AR measurements of the right eye at 2.50 D accommodation.
- The Bland-Altman method was used to estimate intraexaminer repeatability (coefficient of repeatability, COR) and agreement (coefficient of agreement, COA).
- Mean differences (MD) and 95% limits of agreement were calculated for each method.
Main Results:
- Nott retinoscopy (-0.10 +/- 0.66 D) and BCC (-0.05 +/- 0.75 D) demonstrated the best intraexaminer repeatability with the lowest mean differences and narrowest limits of agreement.
- Overall limits of agreement across all methods ranged from +/- 0.92 to +/- 1.00 D, considered clinically significant.
- Dynamic retinoscopy techniques (Nott and MEM) showed better agreement (COA = +/-0.64 D), but MEM exhibited low repeatability (COR = +/-0.98 D).
Conclusions:
- Nott retinoscopy is the most suitable method for clinical assessment of AR due to its high repeatability.
- The BCC technique is also a repeatable and recommended method for clinical practice.
- Despite good agreement between Nott and MEM retinoscopy, poor agreement among other methods precludes their interchangeable clinical use.