Computerized Dynamic Posturography for Postural Control Assessment in Patients with Intermittent Claudication
Published on: December 11, 2013
Accuracy of ankle-brachial index using an automatic blood pressure device to detect peripheral artery disease in
Daniel Benchimol1, Xavier Pillois, Alain Benchimol
1Inserm U828, université Victor-Segalen de Bordeaux 2, hôpital Cardiologique, avenue du Haut-Lévêque, 33604 Pessac, France.
Background:
Previously, we validated determination of ankle-brachial index using an automatic blood pressure device.
Aim:
To test the feasibility and accuracy of the automatic method in assessing pathological ankle-brachial indexes in routine preventive examinations.
Methods:
Two physicians enrolled 354 subjects (74% men) randomly for automatic ankle-brachial index measurements using an OMRON HM 722 device. Ankle-brachial index was calculated by dividing the highest value obtained at each ankle by the highest arm value. Each subject with an abnormal (less than 0.90) automatic index and the six subsequent subjects underwent Doppler index determination.
Results:
Automatic ankle-brachial index determination was possible in both ankles in 350 subjects (99%; mean time 8.1+/-2.1 minutes). The incidence of abnormal automatic ankle-brachial index was 8% (n=28). Correlations between the automatic and Doppler methods were good in left and right legs (r=0.84 and 0.78, respectively; p<0.001). In subjects with an abnormal automatic index, correlations with Doppler indexes were good in both legs (r=0.67, p<0.001). In terms of detecting an abnormal index in a routine preventive examination, the automatic method had good sensitivity (92%), specificity (98%), positive predictive value (86%), negative predictive value (99%) and accuracy (97% compared with the Doppler method). Good results were obtained in subjects with an abnormal index in terms of agreement and concordance with the Doppler method (kappa=0.87).
Conclusions:
The use of a commercially available automatic blood pressure device to detect peripheral artery disease appears feasible and quick in routine medical examinations.
Related Concept Videos
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...