Related Experiment Video
Updated: Jun 21, 2026

06:52
An Inertial Measurement Unit Based Method to Estimate Hip and Knee Joint Kinematics in Team Sport Athletes on the Field
Published on: May 26, 2020
Functional calibration procedure for 3D knee joint angle description using inertial sensors.
J Favre1, R Aissaoui, B M Jolles
1Ecole Polytechnique Fédérale de Lausanne (EPFL), Laboratory of Movement Analysis and Measurement, Lausanne, Switzerland. julien.favre@epfl.ch
Journal of Biomechanics
|August 12, 2009
Summary
A new calibration method using inertial measurement units (IMUs) allows for accurate 3D knee joint angle measurement outside the lab. This ambulatory system offers a promising tool for clinical assessments and therapeutic comparisons.
Area of Science:
- Biomechanics
- Kinesiology
- Medical Instrumentation
Background:
- Accurate 3D knee joint angle measurement is crucial for clinical evaluation and treatment comparison.
- Existing motion capture systems are often lab-bound, necessitating ambulatory solutions for routine practice.
- Inertial Measurement Units (IMUs) show potential for unconstrained knee joint orientation tracking.
Purpose of the Study:
- To develop and validate a novel calibration procedure for IMU-based 3D knee joint angle measurement.
- To adapt the calibration for the joint coordinate system (JCS) using only IMU data.
- To assess the repeatability and accuracy of the proposed method during gait.
Main Methods:
- A new functional calibration procedure utilizing active and passive movements was developed.
- Eight healthy subjects participated in the study.
- The system's performance was evaluated against a reference system during gait, assessing repeatability and measurement errors.
Main Results:
- The calibration procedure demonstrated high repeatability (offset < 1°, SD < 0.3°, CMC > 0.9).
- The system achieved high precision (SD < 2°, CC > 0.75) and moderate accuracy (4.0°–8.1°) for 3D knee angles compared to the reference.
- The IMU-based system combined with the functional calibration is a promising tool for ambulatory knee motion analysis.
Conclusions:
- The proposed IMU-based system and functional calibration offer a reliable method for measuring 3D knee joint angles in ambulatory settings.
- This approach provides a practical tool for clinical examination and monitoring therapeutic interventions.
- The methodology is adaptable for assessing other complex joints like the ankle and elbow.