Related Experiment Video
Updated: Jun 21, 2026

Quantification of Proteins Using Peptide Immunoaffinity Enrichment Coupled with Mass Spectrometry
Published on: July 31, 2011
Application of amino acid analysis using hydrophilic interaction liquid chromatography coupled with isotope dilution
Megumi Kato1, Hisashi Kato, Sakae Eyama
1Bio-Medical Standards Section, Organic Analytical Chemistry Division, National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, 1-1-1 Umezono, Tsukuba, Ibaraki 305-8563, Japan. katou-megu@aist.go.jp
Abstract:
Amino acid analysis that is based on the use of hydrophilic interaction liquid chromatography (HILIC) coupled with isotope dilution mass spectrometry (IDMS) has been developed for the accurate quantification of underivatized amino acids from hydrolyzed protein/peptide. Sufficient separation of amino acids on a zwitterion chromatography (ZIC)-HILIC column was achieved after removal of chloride ions in the hydrolyzate. The detection limits and quantification limits as concentration of the four amino acids ranged from 0.003 to 0.04pmol microL(-1) and from 0.01 to 0.1pmol microL(-1), respectively. The analytical results for the certified reference materials, angiotensin I and bovine serum albumin (BSA), were satisfactory. Furthermore, the quantitative results by this method were compared with those by the commercially available precolumn method, derivatizd with aminoquinolylhydroxysuccinimidyl carbamate (AQC method), and better recovery and more precise data were obtained with this method.