Related Experiment Video
Updated: Jun 21, 2026

Experimental Research Examining How People Can Cope with Uncertainty Through Soft Haptic Sensations
Published on: September 16, 2015
A prospective experience with the lead integrity alert: new certainties and new uncertainties
Thorben Koenig1, Ajmal Gardiwal, Hanno Oswald
1Department of Cardiovascular Medicine, Hannover Medical School, D-30625 Hannover, Germany. koenig.thorben@mh-hannover.de
Abstract:
Lead fracture is one of the major complications in implantable cardioverter defibrillator (ICD) therapy often leading to a series of inadequate shocks and thus greatly impairing quality of life of patients. The novel lead integrity alert algorithm by Medtronic (Medtronic Inc., Minneapolis, MN, USA) is addressing this problem. We report a case of a lead failure being correctly predicted by shifting lead impedances about 4 weeks before the first episode of oversensing. Additionally, our case illustrates the new problem of how to clinically handle a patient with a highly probable, but not completely certain, ICD-lead failure.