A validation of electrohysterography for uterine activity monitoring during labour
Benoit C Jacod1, Elisabeth Margo Graatsma, Eline Van Hagen
1Department of Obstetrics, University Medical Centre, Utrecht, The Netherlands. bjacod@gmail.com
Objectives:
Validation of electrohysterography (EHG) as a method for uterine activity monitoring during labour by comparing with intra-uterine pressure catheter (IUPC) recordings.
Design:
Prospective observational study.
Setting:
Labour ward in a tertiary centre in the Netherlands.
Population:
Thirty-two women in labour.
Methods:
Simultaneous recording of uterine activity with EHG and IUPC for at least 30 min.
Main Outcome Measures:
Number of uterine contractions detected by both EHG and IUPC (sensitivity). Number of contractions detected by EHG only [positive predictive value (PPV)]. Correlation between contraction amplitude and duration measured by EHG and IUPC.
Results:
EHG detects uterine contractions accurately: sensitivity = 94.5% (95%CI: 87.5-100), PPV = 88.3% (95%CI: 76.2-100). The correlation of contractions' duration and amplitude between both methods is r = 0.31 (95%CI: 0.23-0.39) and r = 0.45 (95%CI: 0.38-0.52), respectively.
Conclusions:
EHG detects uterine contractions accurately during labour but the contraction's characteristics it measures are not directly comparable with that of IUPC.