Intraexaminer and interexaminer reliability of mastoid fossa readings using a temporal artery thermometer
Claudia Seay1, Cindy Gibbon, John Hart
1Associate Professor, Clinical Sciences, Sherman College of Straight Chiropractic, P.O. Box 1452, Spartanburg, SC 29304.
Objective:
The temperature of the mastoid fossa region has been measured by chiropractors since the 1950s using various instruments. The reliability of the procedure is largely unknown for the various instruments used for this purpose. This study assessed the reliability of a thermal instrument designed to measure forehead temperature while having an alternate landmark as the mastoid fossa.
Methods:
: Thirty students were recruited to participate. Three blinded examiners twice scanned each student with the temporal artery thermometer. Intraexaminer and interexaminer reliability was assessed using the intraclass correlation coefficient. Differential margins of error were also calculated.
Results:
Intraexaminer reliability was acceptable in 6 of 6 assessments, whereas interexaminer reliability was acceptable in 3 of 4 assessments. The maximum mean differential margin of error was 0.54 for intraexaminer reliability and 0.46 for interexaminer reliability.
Discussion:
Combining the mean maximum differential margin of error for intraexaminer (0.54 degrees ) and interexaminer (0.46 degrees ) and then averaging these 2 ([0.54 + 0.46] / 2) results in a maximum mean differential margin of error of 0.50 degrees .
Conclusion:
In this study, intraexaminer reliability ranged from fair to strong. For these examiners, fossa differentials of 0.5 can be considered within the margin of error.
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...