Automated Acoustic Dispensing for the Serial Dilution of Peptide Agonists in Potency Determination Assays
Published on: November 10, 2016
Establishing quality assurance criteria for serial dilution operations on liquid-handling equipment
Ioana Popa-Burke1, Brian Lupotsky, Joseph Boyer
1GlaxoSmithKline, Sample Management Technologies, Research Triangle Park, North Carolina 27709-3398, USA. ioana.g.popa-burke@gsk.com
Abstract:
Since the advent of high-throughput screening (HTS) in the early 1990s, parallel multichannel liquid handlers have become a mainstay in every drug discovery setting. Although several peer-reviewed publications have discussed methods and criteria for stamping multiwell copies, there is very little information about establishing a standard operating procedure (SOP) for standard (microliter-level) serial dilutions of compounds used in dose-response experiments. The authors discuss the 4 main criteria any serial dilution process must pass (accuracy, precision, fold dilution, and outliers) and the process for establishing thresholds for all of these values in a compound management or biological screening laboratory. The thresholds need to be both low enough to be acceptable from a biological potency variability perspective and high enough to allow the instruments to pass the quality assurance (QA) analysis on a regular basis. In this article, the authors suggest suitable thresholds arrived at by a variety of methods, including trend analysis of QA data, survey questionnaire from the main stakeholders (screening scientists, chemists), and published criteria for single-shot stamping. A mathematical analysis of the effect of threshold values on estimated XC(50)s was performed to ensure that the variability introduced by the serial dilution step is within acceptable overall variability limits.
Related Concept Videos
Scale-Up Processes
Data Validation
Key parameters for method validation include:
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Quality Assurance
Drug Dissolution: Requirements and Profile Comparison
Standard Solutions