ATP measurement as method to monitor the quality of reprocessing flexible endoscopes

Dorothea Hansen1, Daniel Benner, Martin Hilgenhöner

  • 1Krankenhaushygiene, Universitätsklinikum Essen, Germany. dorothea.hansen@uni-essen.de

Insights

Adenosine triphosphate (ATP) bioluminescence offers a rapid method for checking flexible endoscope reprocessing hygiene. While not replacing microbiological cultures, it can quickly flag scopes needing immediate inspection to reduce patient infection risk.

Area of Science:

  • Microbiology
  • Infection Control
  • Medical Device Reprocessing

Background:

  • Inadequate cleaning of flexible endoscopes poses a significant patient infection risk.
  • Current quality control relies on microbiological cultures with delayed results.
  • Rapid methods for assessing endoscope reprocessing are needed.

Purpose of the Study:

  • To compare the efficacy of adenosine triphosphate (ATP) bioluminescence with routine microbiological cultures for evaluating flexible endoscope reprocessing.
  • To determine the sensitivity and specificity of ATP bioluminescence as a hygiene indicator.

Main Methods:

  • 108 flexible endoscopes underwent reprocessing and subsequent testing.
  • ATP bioluminescence assays were performed for rapid hygiene assessment.
  • Results were compared against traditional microbiological culture techniques.

Main Results:

  • Bacterial growth was detected in 28 endoscopes via microbiological cultures.
  • ATP bioluminescence identified between 28 and 67 endoscopes as positive, depending on the threshold used.
  • Sensitivity ranged from 0.46 to 0.75, and specificity from 0.43 to 0.81.

Conclusions:

  • ATP bioluminescence is not a substitute for standard microbiological testing.
  • ATP bioluminescence can serve as a valuable, rapid tool to indicate the need for immediate reprocessing checks.
  • Implementing ATP bioluminescence can potentially enhance patient safety by enabling quicker interventions.