Fluorescence intensity calibration using the Raman scatter peak of water
Related Experiment Video
Updated: Jun 21, 2026

A Multimodal Wide-Field Fourier-Transform Raman Microscope
Published on: December 30, 2025
1Dept. Food Science, Faculty of Life Sciences, University Copenhagen, Rolighedsvej 30, DK-1958, Frederiksberg, Denmark. ajla@life.ku.dk
Abstract:
Fluorescence data of replicate samples obtained from different fluorescence spectrometers or by the same spectrometer but with different instrument settings can have great intensity differences. In order to compare such data an intensity calibration must be applied. Here we explain a simple calibration method for fluorescence intensity using only the integrated area of a water Raman peak. By applying this method to data from three different instruments, we show that it is possible to remove instrument-dependent intensity factors, and we present results on a unified scale of Raman units. The method presented is a rapid and simple approach suitable for routine measurements with no need for hazardous chemicals.
Related Concept Videos
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...