Transient local injury current in right ventricular electrogram after implantable cardioverter-defibrillator shock
Larisa G Tereshchenko1, Mitchell N Faddis, Barry J Fetics
1Division of Cardiology, Department of Medicine, Johns Hopkins Hospital, Baltimore, Maryland 21287, USA. lteresh1@jhmi.edu
Objectives:
This study aimed to identify an early marker of functional impairment after an implantable cardioverter-defibrillator (ICD) shock as a predictor of heart failure progression.
Background:
The ICD population has substantial risk of death due to progressive pump failure.
Methods:
Near-field (NF) bipolar right ventricular (RV) electrograms (EGMs) during induced ventricular fibrillation (VF) and 10 s after rescue ICD shock were analyzed in 310 patients (mean age 59 +/- 14.5 years, 219 men [71%]) with structural heart disease, New York Heart Association functional class I to III, and implanted with a single- or dual-chamber Medtronic (Minneapolis, Minnesota) ICD for primary (245 patients, 79%) or secondary prevention of sudden cardiac arrest. A local injury current (LIC) on NF RV EGM was defined as a deviation of EGM potential > or =1 mV or > or =15% of the preceding R-wave peak-to-peak amplitude.
Results:
During mean follow-up of 29.3 +/- 15.0 months, the combined end point of death or hospitalization due to congestive heart failure (CHF) exacerbation was documented in 40 patients (12.9%, or 5.3% per person-year of follow-up). LIC was observed in 106 patients. In multivariate risk analysis, after adjustment for baseline prognostic factors (ejection fraction, history of atrial fibrillation, diabetes mellitus) and appropriate ICD shocks during follow-up, patients with observed LIC after induced VF rescue ICD shock at ICD implantation were more likely to die or to be hospitalized (hazard ratio: 2.69; 95% confidence interval: 1.41 to 5.14; p = 0.003).
Conclusions:
Transient LIC on bipolar NF RV EGM after induced VF rescue ICD shock is associated with increased risk of CHF progression, future hospitalizations due to CHF exacerbation, and subsequent heart failure death.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Heart Failure IV: Classification and Diagnostic Evaluation