The accuracy of central SBP determined from the second systolic peak of the peripheral pressure waveform
Stacey S Hickson1, Mark Butlin, Fraz A Mir
1University of Cambridge, Addenbrooke's Hospital, UK. sh529@cam.ac.uk
Background:
Recent evidence suggests that central aortic blood pressure may be a better predictor of cardiovascular risk than peripheral blood pressure. The central SBP (cSBP) can be estimated from the late systolic shoulder of the radial pulse waveform. We compared the second systolic peak of the radial waveform (pSBP(2)) with the central systolic pressure derived by a generalized transfer function in a large cohort, across a wide age range, of patients from the Anglo-Cardiff Collaborative Trial. We also compared pSBP(2) with the true cSBP measured by cardiac catheterization [invasively measured cSBP (cSBPi)].
Methods:
Noninvasive measurements were made by applanation tonometry using the SphygmoCor device. The aortic pressure waveform was derived from the radial waveform using a validated transfer function. Invasive measures of cSBPi were carried out in a group of 38 patients undergoing diagnostic cardiac angiography, and radial artery pressure waveforms were simultaneously recorded using the SphygmoCor device.
Results:
Overall, there was a strong correlation (r = 0.99, P<0.001) and good agreement between pSBP(2) and the derived cSBP (mean difference +/- SD 1 +/- 4 mmHg). However, there was a systematic bias with a greater difference between these measures at lower average pressures. There was also a strong correlation and good agreement between the invasively measured cSBPi and pSBP(2) (r = 0.92, P<0.001, mean difference 2 +/- 6 mmHg).
Conclusion:
The second systolic shoulder of the peripheral pressure waveform approximates the cSBP in a large cohort of patients across a wide age range, but this may be inaccurate at low SBP values.
More Related Videos
06:04Pulse-Wave Velocity, Flow-Mediated Dilation, and Carotid Intima-Media Thickness to Assess Cardiovascular Risk in Population with Metabolic Syndrome
Published on: September 27, 2024
05:51Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
Published on: May 3, 2018
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation