Validation of the PhysioQuant blood pressure measuring monitor in adults
Wolf Langewitz1, Sabine Tanner
1Department of Psychosomatic Medicine, Internal Medicine, University Hospital Basel, Basel, Switzerland. wlangewitz@uhbs.ch
Abstract:
A new automatic blood pressure (BP) measuring device PhysioQuant (also marketed under the brand name Ergoscan) was evaluated according to the International Protocol for Validation of Blood Pressure Measuring Devices in adults issued by the European Society of Hypertension. BP values measured by the PhysioQuant were compared with BP readings from two independent observers. Thirty-three patients (13 male, 20 female) provided systolic and diastolic BP readings in the normotensive, borderline hypertensive, and hypertensive range. Their age varied between 29 and 93 years, their arm circumference between 23.5 and 37 cm. The device showed a mean (+/-SD) deviation from observer measurements of -0.1 (+/-3.4) mmHg for systolic and +1.1 (+/-3.9) mmHg for diastolic BP. The accuracy of the device did not vary according to BP values or other patient characteristics. The device passed all phases of the protocol and can be recommended following the regulation rules of the European Society of Hypertension.
Related Concept Videos
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure