Quantitative three-dimensional photoacoustic tomography of the finger joints: phantom studies in a spherical scanning
1Department of Biomedical Engineering, University of Florida, Gainesville, FL 32611, USA.
Abstract:
In this paper, we present a series of phantom experiments that aim to systematically evaluate the possibility of quantitative three-dimensional (3D) photoacoustic imaging of the finger joints in a spherical scanning geometry. The 3D absorption coefficient images of the joint-mimicking phantoms are obtained using our finite-element-based photoacoustic image reconstruction algorithm coupled with the photon diffusion equation. The results show that a 1 mm thick 'cartilage' can be accurately differentiated from the 'bones' with a 1 MHz transducer. In addition, the absorption coefficient of the 'cartilage' can be effectively recovered when this optical property is varied from 0.015 to 0.04 mm(-1). The phantom images shown in this paper suggest that 3D photoacoustic tomography in a spherical scanning geometry has the potential to image a realistic human joint.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...