Automated pressure-controlled discography with constant injection speed and real-time pressure measurement
1Department of Neurosurgery, Presbyterian Medical Center, Jeonju, Korea.
Objective:
This study was designed to investigate automated pressure-controlled discography (APCD) findings, to calculate the elastance of intervertebral discs, and to assess the relationship between the calculated elastance and disc degeneration.
Methods:
APCD was performed in 19 patients. There were a total of 49 intervertebral discs treated. Following intradiscal puncture, a dye was constantly injected and the intradiscal pressure was continuously measured. The elastance of the intervertebral disc was defined as unit change in intradiscal pressure per fractional change in injected dye volume. Disc degeneration was graded using a modified Dallas discogram scale.
Results:
The mean elastance was 43.0 +/- 9.6 psi/mL in Grade 0, 39.5 +/- 8.3 psi/mL in Grade 1, 30.5 +/- 22.3 psi/mL in Grade 2, 30.5 +/- 22.3 psi/mL in Grade 3, 13.2 +/- 8.3 psi/mL in Grade 4 and 6.9 +/- 3.8 psi/mL in Grade 5. The elastance showed significant negative correlation with the degree of degeneration (R(2) = 0.529, p = 0.000).
Conclusion:
APCD liberates the examiner from the data acquisition process during discography. This will likely improve the quality of data and the reliability of discography. Elastance could be used as an indicator of disc degeneration.
More Related Videos
06:00Wet Beveling of Microinjection Needles Utilizing Constant Air Pressure for Feedback on Needle Opening
Published on: September 27, 2024
09:46A Pressure Injection System for Investigating the Neuropharmacology of Information Processing in Awake Behaving Macaque Monkey Cortex
Published on: March 14, 2016
Related Concept Videos
Bioreactor Controls-I
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...