Related Experiment Video
Updated: Jun 20, 2026

Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
Use of cardiac output to improve measurement of input function in quantitative dynamic contrast-enhanced MRI
Jeff L Zhang1, Henry Rusinek, Louisa Bokacheva
1Department of Radiology, New York University School of Medicine, New York, NY, USA. Lei.Zhang@nyumc.org
Purpose:
To validate a new method for converting MR arterial signal intensity versus time curves to arterial input functions (AIFs).
Materials And Methods:
The method constrains AIF with patient's cardiac output (Q). Monte Carlo simulations of MR renography and tumor perfusion protocols were carried out for comparison with two alternative methods: direct measurement and population-averaged input function. MR renography was performed to assess the method's inter- and intraday reproducibility for renal parameters.
Results:
In simulations of tumor perfusion, the precision of the parameters (K(trans) and v(e)) computed using the proposed method was improved by at least a factor of three compared to direct measurement. Similar improvements were obtained in simulations of MR renography. Volunteer study for testing interday reproducibility confirmed the improvement of precision in renal parameters when using the proposed method compared to conventional methods. In another patient study (two injections within one session), the proposed method significantly increased the correlation coefficient (R) between GFR of the two exams (0.92 vs. 0.83) compared to direct measurement.
Conclusion:
A new method significantly improves the precision of dynamic contrast-enhanced (DCE) parameters. The method may be especially useful for analyzing repeated DCE examinations, such as monitoring tumor therapy or angiotensin converting enzyme-inhibitor renography.
Related Concept Videos
Magnetic Resonance Imaging
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Imaging Studies for Cardiovascular System V: CT