Focusing, collimation and flux throughput at the IMCA-CAT bending-magnet beamline at the Advanced Photon Source
Irina Koshelev1, Rong Huang, Timothy Graber
1IMCA-CAT, The Center for Advanced Radiation Sources, University of Chicago, IL 60637, USA. ikoshelev@anl.gov
Abstract:
The IMCA-CAT bending-magnet beamline was upgraded with a collimating mirror in order to achieve the energy resolution required to conduct high-quality multi- and single-wavelength anomalous diffraction (MAD/SAD) experiments without sacrificing beamline flux throughput. Following the upgrade, the bending-magnet beamline achieves a flux of 8 x 10(11) photons s(-1) at 1 A wavelength, at a beamline aperture of 1.5 mrad (horizontal) x 86 microrad (vertical), with energy resolution (limited mostly by the intrinsic resolution of the monochromator optics) deltaE/E = 1.5 x 10(-4) (at 10 kV). The beamline operates in a dynamic range of 7.5-17.5 keV and delivers to the sample focused beam of size (FWHM) 240 microm (horizontally) x 160 microm (vertically). The performance of the 17-BM beamline optics and its deviation from ideally shaped optics is evaluated in the context of the requirements imposed by the needs of protein crystallography experiments. An assessment of flux losses is given in relation to the (geometric) properties of major beamline components.
More Related Videos
14:11Quantification of Hydrogen Concentrations in Surface and Interface Layers and Bulk Materials through Depth Profiling with Nuclear Reaction Analysis
Published on: March 29, 2016
07:11Fully Autonomous Characterization and Data Collection from Crystals of Biological Macromolecules
Published on: March 22, 2019