Reprocessing and sterilization of single-use electrophysiological catheters: removal of organic carbon and protein
Bruce R Lester1, Nicole P Boser, Keith Miller
1SterilMed, Inc., R&D Division, 11400 73rd Ave North, Maple Grove, MN 55369, USA. blester@sterilmed.com
Abstract:
Although evidence to date indicates that reprocessing electrophysiological (EP) catheters results in clean, sterile devices, some concerns persist with regard to the risk of residual contamination. We examined the ability of a defined reprocessing procedure coupled with a validated sterilization protocol to remove organic carbon and protein residues from worst-case soiled EP catheters resulting in clean, sterile devices. Total organic carbon (TOC) determinations indicated that detergent residues on reprocessed used catheters were nominal and significantly lower than organic carbon levels present on new catheters. Determination of the mean residual organic carbon and protein contaminants on soiled and reprocessed EP catheters further indicated that TOC and protein were reduced (> or =99% of residue removed) below previously reported levels and current accepted standards. Moreover, reprocessed end-of-life catheters (six clinical uses, plus five reprocessings) were examined for residual microorganisms and found to be sterile. End-of-life catheters that had been inoculated with >10(6) CFU Bacillus atrophaeus spores and subjected to a half-cycle ethylene oxide exposure were also found to be sterile. Our data indicate that EP catheter cleaning and reprocessing using defined protocols effectively removes detergent residues and biological contamination, and provides sterile devices.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Downstream Processing