Rapid chairside sterilization of endodontic files using 6% sodium hypochlorite
Heather L Gnau1, Gary G Goodell, Glen M Imamura
1Naval Postgraduate Dental School, Bethesda, MD, USA.
Introduction:
The purpose of this study was to assess the percentage of new endodontic files (taken directly from the manufacturers' packages) that were contaminated with viable microorganisms and to determine the amount of time new files needed to be immersed in 6% sodium hypochlorite to achieve sterility.
Methods:
Endodontic files from four manufacturers were placed either immediately in thioglycolate broth or immersed in 6% sodium hypochlorite for 1, 2, or 5 minutes before placement in the broth. The files were aerobically incubated for 72 hours and sterility determined by the absence of broth turbidity, as determined by a blinded examiner. The data were analyzed using the Cochran-Q test followed by the McNemar test for pair-wise comparisons. The level of statistical significance was set at alpha = 0.05.
Results:
When pooling all manufacturers' files, no significant differences were found between groups.
Conclusion:
New files demonstrated a 6% contamination rate and none of the immersion times in NaOCl achieved file sterility.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Hand hygiene
Hand washing...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Chemical Agents for Microbial Control