Related Experiment Video
Updated: Jun 20, 2026

Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
Ex vivo accuracy of an apex locator using digital signal processing in primary teeth
Mário Roberto Leonardo1, Lea Assed Bezerra da Silva, Paulo Nelson-Filho
1Department of Pediatric Clinic, Preventive and Community Dentistry, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, São Paulo, Brazil.
Purpose:
The purpose of this study was to evaluate ex vivo the accuracy an electronic apex locator during root canal length determination in primary molars.
Methods:
One calibrated examiner determined the root canal length in 15 primary molars (total=34 root canals) with different stages of root resorption. Root canal length was measured both visually with the placement of a K-file 1 mm short of the apical foramen or the apical resorption bevel, and electronically using an electronic apex locator (Digital Signal Processing). Data were analyzed statistically using the intraclass correlation (ICC) test.
Results:
Comparing the actual and electronic root canal length measurements in the primary teeth showed a high correlation (ICC=0.95).
Conclusions:
The Digital Signal Processing apex locator is useful and accurate for apex foramen location during root canal length measurement in primary molars.
Related Concept Videos
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Tooth Anatomy
The Crown, Neck, and Root
The visible part of the tooth is referred to as the crown. It's covered by enamel, the hardest substance in the human body. The crown is uniquely shaped for each type of tooth, allowing for different functions such as cutting, tearing, or grinding food.