Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol
07:59

Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol

Published on: September 7, 2018

A new phantom using polyethylene glycol as an apparent diffusion coefficient standard for MR imaging.

Ryohei Matsuya1, Masahiro Kuroda, Yoshitsugu Matsumoto

  • 1Faculty of Health Sciences, Graduate School of Health Sciences, Okayama University, Okayama 700-8558, Japan.

International Journal of Oncology
|September 3, 2009
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Post-mastectomy radiation therapy in Poland syndrome with left breast cancer: A case report.

Experimental and therapeutic medicine·2026
Same author

Microstructural changes in irradiated teeth revealed by swept-source optical coherence tomography.

Scientific reports·2026
Same author

Analgesic effect of robotic stroking with speech.

Scientific reports·2026
Same author

Lessons from the past: what a national e-prescribing protocol tells us about the limits of restrictive approaches to antimicrobial stewardship.

Clinical microbiology and infection : the official publication of the European Society of Clinical Microbiology and Infectious Diseases·2026
Same author

Oral Health Status in Older Care Recipients and Family Caregiver Perceptions of Caregiving Burden.

Gerodontology·2026
Same author

A Novel <i>Paludibacterium</i> Species Isolated from Human Blood.

Microorganisms·2026

Researchers created a new, low-cost phantom using polyethylene glycol to help calibrate MRI scanners. This tool allows clinicians to accurately mimic the diffusion properties of various human tissues, improving the reliability of diffusion-weighted imaging for detecting tumors and strokes.

Area of Science:

  • Medical imaging physics and the development of an Apparent Diffusion Coefficient standard
  • Diagnostic radiology instrumentation within clinical MRI research

Background:

No prior work had resolved the challenge of creating a versatile reference tool for diffusion-weighted imaging. Current calibration devices often rely on limited material types and restricted temperature ranges for validation. This uncertainty drove the need for a more flexible solution in clinical settings. Prior research has shown that existing standards fail to provide a wide range of diffusion values. That limitation hinders the development of new pulse sequences for whole-body screening. This gap motivated the exploration of alternative substances for phantom construction. Scientists previously struggled to predict how material concentration and temperature interact to influence diffusion measurements. Establishing a reliable standard remains a priority for improving diagnostic accuracy in magnetic resonance imaging.

Purpose Of The Study:

The aim of this study was to develop a new phantom material that provides arbitrary diffusion values for calibrating imaging equipment. Researchers sought to overcome the limitations of existing standards, which often lack flexibility in concentration and temperature ranges. The team investigated whether polyethylene glycol could serve as a reliable substance for this purpose. They intended to create a mathematical framework that allows for the precise calculation of material properties at any given temperature. This motivation stemmed from the difficulty of predicting diffusion behavior in previous experimental setups. By developing empirical formulas, the authors hoped to simplify the creation of customized phantoms. They focused on ensuring that these tools could mimic the characteristics of various human tissues. This work addresses the need for a versatile, cost-effective, and safe standard for diffusion-weighted imaging applications.

Keywords:
diffusion-weighted imagingphantom calibrationpolyethylene glycoldiagnostic radiology

Frequently Asked Questions

The researchers propose using polyethylene glycol to create phantoms with specific diffusion properties. By adjusting the concentration of this polymer and the ambient temperature, they can achieve a wide range of values, unlike previous methods that relied on fixed, limited material compositions.

The team utilized seven distinct concentrations of polyethylene glycol to calibrate their mathematical models. This range allowed them to map the relationship between polymer density, thermal conditions, and the resulting diffusion measurements, providing a comprehensive dataset for their empirical formulas.

The authors state that precise temperature control is necessary because diffusion rates are highly sensitive to thermal fluctuations. By incorporating temperature variables into their formulas, they ensure the phantom remains accurate across the 18 to 45 degrees Celsius range.

More Related Videos

Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents
08:59

Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents

Published on: April 15, 2016

Biofunctionalized Prussian Blue Nanoparticles for Multimodal Molecular Imaging Applications
11:28

Biofunctionalized Prussian Blue Nanoparticles for Multimodal Molecular Imaging Applications

Published on: April 28, 2015

Related Experiment Videos

Last Updated: Jun 20, 2026

Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol
07:59

Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol

Published on: September 7, 2018

Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents
08:59

Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents

Published on: April 15, 2016

Biofunctionalized Prussian Blue Nanoparticles for Multimodal Molecular Imaging Applications
11:28

Biofunctionalized Prussian Blue Nanoparticles for Multimodal Molecular Imaging Applications

Published on: April 28, 2015

Main Methods:

The review approach involved developing empirical formulas to relate polymer concentration and thermal conditions to diffusion properties. Investigators prepared seven distinct solutions using varying amounts of the chosen material. These samples underwent systematic scanning while the ambient environment was gradually heated from 17 to 46 degrees Celsius. Researchers recorded data at one-degree intervals to ensure high-resolution mapping of the physical properties. The team then processed these images to derive two specific mathematical models. One formula calculates diffusion values based on concentration and temperature, while the other determines the required concentration for a target value. Validation occurred by comparing these calculated predictions against measurements from newly constructed samples. This rigorous testing confirmed the reliability of the predictive framework across the specified operational parameters.

Main Results:

The strongest finding confirms that the empirical formulas provide high accuracy when compared to measured diffusion values. The researchers successfully established a range for diffusion coefficients from 0.37 to 3.67 times ten to the negative third power square millimeters per second. Their formulas remain valid for polymer concentrations spanning 0 to 120 millimolar. The team also verified that the models function reliably between 18 and 45 degrees Celsius. Comparisons between predicted and observed data demonstrated minimal variance in the results. This consistency supports the utility of the phantom for mimicking various clinical lesions. The study highlights that the materials used are inexpensive and simple to assemble. These results provide a clear pathway for creating standardized references in diagnostic imaging.

Conclusions:

The authors propose that their polyethylene glycol phantoms offer a practical and cost-effective alternative to existing calibration tools. Synthesis and implications suggest that these materials are safe and straightforward to prepare for routine laboratory use. The researchers demonstrate that their mathematical models accurately predict diffusion values across a broad range of conditions. This work indicates that clinicians can now generate custom phantoms to match specific tissue characteristics. The findings imply that these formulas support the creation of standards for diverse clinical lesions. The study confirms that the proposed method maintains high precision within defined temperature and concentration limits. The authors suggest that this approach enhances the standardization of diffusion-weighted imaging protocols. These results provide a robust framework for future quality control in diagnostic imaging environments.

The researchers employed diffusion-weighted imaging data to derive their predictive models. This imaging modality provides the raw signal intensity required to calculate the diffusion coefficients, which are then used to validate the accuracy of the newly developed concentration formulas.

The study measured diffusion values ranging from 0.37 to 3.67 times ten to the negative third power square millimeters per second. This range covers the typical diffusion characteristics observed in various human clinical lesions, making the phantom highly relevant for diagnostic applications.

The authors suggest that these phantoms could become a standard tool for calibrating magnetic resonance imaging scanners. They propose that this method will facilitate the development of new pulse sequences by providing a reliable, reproducible reference for diverse clinical scenarios.