Effects of heterogeneous electron-transfer rate on the resolution of electrophoretic separations based on
Joseph Wang1, Baomin Tian, Madhu Prakash Chatrathi
1Department of Nanoengineering, University of California San Diego, La Jolla, CA 92093, USA. josephwang@ucsd.edu
Abstract:
We demonstrate here that the electrode kinetics of an electrochemical detector contributes greatly to the resolution of the analyte bands in microchip electrophoresis systems with amperometric detection. The separation performance in terms of resolution and theoretical plate number can be improved and tailored by selecting or modifying the working electrode and/or by controlling the detection potential. Such improvements in the separation performance reflect the influence of the heterogeneous electron-transfer rate of electroactive analytes upon the post-channel band broadening, as illustrated for catechol and hydrazine compounds. The electrode kinetics thus has a profound effect not only on the sensitivity of electrochemical detectors but on the separation efficiency and the overall performance of microchip electrochemistry systems.
More Related Videos
13:42Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
05:45Capillary Electrophoresis-based Hydrogen/Deuterium Exchange for Conformational Characterization of Proteins with Top-down Mass Spectrometry
Published on: June 8, 2021
Related Concept Videos
Electrophoresis: Overview
There...
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Ion-Exchange Chromatography
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...