A pilot, randomized, double-blind, cross-over study of high cut-off versus high-flux dialysis membranes
Darren Lee1, Michael Haase, Anja Haase-Fielitz
1Department of Nephrology, Austin Hospital, University of Melbourne, Melbourne, Vic., Australia.
Background:
High cut-off (HCO) membranes may increase beta(2)-microglobulin (beta2M) removal compared to standard high-flux membranes.
Methods:
Eight stable haemodialysis patients were enrolled in a prospective, randomized, double-blind, cross-over study and treated with HCO and high-flux membranes for 2 weeks each, between a 1-week washout period. Primary end point was serum beta2M removal. Secondary end points included serum albumin concentrations, albumin and small solute clearances.
Results:
HCO membranes achieved significantly lower median post-dialysis beta2M concentration (10.8 vs. 14.2 mg/l; p = 0.003) and greater beta2M reduction ratio (62.3 vs. 51.0%; p < 0.002). Serum albumin decreased with HCO membranes (from 36 to 29.5 g/l; p = 0.018) but increased to 33.5 g/l after the washout period. Albumin clearance was significantly greater with HCO membranes (2.2 vs. 0.06 ml/min; p = 0.004). Urea reduction ratio was significantly lower with HCO membranes (64.8 vs. 71.5%; p < 0.001).
Conclusion:
beta2M removal was superior with HCO membranes. Reduction in serum albumin and lower small solute clearance require further investigations.
Related Concept Videos
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Hemodialysis II: Procedure and Complications
Hemodialysis I: Introduction
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis