Related Experiment Video
Updated: Jun 20, 2026

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Dosimetric consequences of uncorrected setup errors in helical Tomotherapy treatments of breast-cancer patients
S Murty Goddu1, Sridhar Yaddanapudi, Olga L Pechenaya
1Department of Radiation Oncology, Washington University School of Medicine, 4921 Parkview Place, Campus Box 8224, St. Louis, MO 63110, USA. sgoddu@radonc.wustl.edu
Background And Purpose:
The Tomotherapy Hi-Art II system allows acquisition of pre-treatment MVCT images to correct patient position. This work evaluates the dosimetric impact of uncorrected setup errors in breast-cancer radiation therapy.
Materials And Methods:
Breast-cancer patient-positioning errors were simulated by shifting the patient computed-tomography (CT) dataset relative to the planned photon fluence and re-computing the dose distributions. To properly evaluate the superficial region, film measurements were compared against the Tomotherapy treatment planning system (TPS) calculations. A simulation of the integrated dose distribution was performed to evaluate the setup error impact over the course of treatment.
Results:
Significant dose differences were observed for 11-mm shifts in the anterolateral and 3-mm shifts in the posteromedial directions. The results of film measurements in the superficial region showed that the TPS overestimated the dose by 14% at a 1-mm depth, improving to 3% at depths >or=5mm. Significant dose reductions in PTV were observed in the dose distributions simulated over the course of treatment.
Conclusions:
Tomotherapy's rotational delivery provides sufficient photon fluence extending beyond the skin surface to allow an up to 7-mm uncorrected setup error in the anterolateral direction. However, the steep dose falloff that conforms to the lung surface leads to compromised dose distributions with uncorrected posteromedial shifts. Therefore, daily image guidance and consequent patient repositioning is warranted for breast-cancer patients.