Fentanyl Analog Screening using LC-TIMS-TOF MS/MS
Published on: November 8, 2024
Validated LC coupled to ESI-MS/MS analysis for fentanyl in human plasma and UV analysis in applied reservoir
1Department of Hospital Pharmacy, Hamamatsu University School of Medicine, Higashi-ku, Hamamatsu, Japan.
Background And Objectives:
Fentanyl has been used for cancer pain in transdermal formulation. The aim of the present study was to establish an analytical method for fentanyl in human plasma and in an applied transdermal reservoir patch (Reservoir-TTS), as well as for therapeutic monitoring of fentanyl in cancer patients.
Method:
Electro-spray ionization mass spectrometric (ESI-MS/MS) analysis followed solid phase extraction (SPE) from human plasma and drug reservoir extraction from an applied Reservoir-TTS. Each separation was completed within 9 min using an ODS column (particle size, 3 microm, 2.0 mm i.d. x 75 mm) with 25% acetonitrile containing 5 mm ammonium acetate at pH 3.5. In the ESI-MS/MS analysis, the calibration curve for fentanyl was linear over a concentration range of 0.05-7.2 ng/mL in human plasma. The extraction efficiency of fentanyl in the human plasma was more than 95%. The intra- and interassay precision and accuracy were within 7% and 97.3-101.2%, respectively. The lower LOQ for fentanyl was 0.05 ng/mL in the human plasma. The extraction of the 25 microg/h and 50 microg/h Reservoir-TTS gave reproducible recoveries of 88.3% and 90.9%, respectively. The plasma concentration of fentanyl showed large interindividual variation in 31 patients with cancer pain.
Conclusion:
The method described is simple, accurate, and reproducible, and should be helpful for the therapeutic monitoring of fentanyl in cancer patients.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...