Assessment of radiation doses to cardiologists during interventional examinations
1Department of Medical Physics, Ankara University, Institute of Nuclear Science, 06100 Tandogan, Ankara, Turkey. bor@eng.ankara.edu.tr
Insights
Two methods accurately estimate cardiologist radiation doses during cardiac procedures. One correlates staff extremity doses with patient dose-area product (DAP), while the other uses phantom measurements to calculate scatter doses.
Area of Science:
- Medical Physics
- Radiological Protection
- Cardiology
Background:
- Cardiologists performing interventional cardiac procedures are exposed to ionizing radiation.
- Accurate estimation of occupational radiation doses is crucial for implementing effective radiation protection measures.
Purpose of the Study:
- To investigate the accuracy of two on-line dosimetric techniques for estimating cardiologist radiation doses.
- To establish a database correlating cardiologist extremity doses with patient dose-area product (DAP) values.
- To compare calculated cardiologist doses with measured values using a second method.
Main Methods:
- Method 1: Established a database by measuring doses of nine cardiologists and patient DAP during 166 interventional cardiac examinations across five departments.
- Measured mean doses to eyes, thyroid, waist, wrists, and legs per procedure.
- Calculated effective doses considering protective tools.
- Method 2: Measured scatter doses using a Rando phantom under conditions similar to patient procedures.
- Calculated cardiologist doses based on phantom exposures and patient examination parameters.
Main Results:
- Mean doses per procedure: eyes 72.4 µGy, thyroid 68.5 µGy, waist 11.2 µGy, wrists 67.8–216 µGy, legs 137–384 µGy.
- Mean effective dose: 12.14 µSv/procedure.
- Staff dose to patient DAP ratios ranged from 0.14 to 3.75.
- Second method involved calculating and comparing doses with measured values.
Conclusions:
- Both investigated on-line dosimetric techniques show accuracy in estimating cardiologist radiation doses.
- The established database and scatter dose calculation methods provide valuable tools for occupational radiation monitoring in cardiology.
- Findings support the implementation of robust radiation safety protocols for healthcare professionals in interventional cardiology.
Abstract:
The accuracy of two on line dosimetric techniques was investigated for the estimation of cardiologist doses. The first technique involves the establishment of a database relating the cardiologist extremity doses to patient DAP values. Doses of nine cardiologists were measured together with patient doses during the interventional cardiac examinations of 166 patients for this purpose. Data were collected from five cardiology departments. The mean of the eye, thyroid, waist, right-left wrist, and right-left leg doses per procedure were measured as 72.4 (31.6-107.1), 68.5 (13.3-174.6), 11.2 (0.9-28.4), 67.8 (21.9-120.3) to 216 (52.7-425.4), and 137 (51.4-386.2) to 384 (135-1168.3) microGy/procedure. The effective doses were calculated according to the use of protection tools and a mean value of 12.14 (1.2-30.2) microSv/procedure was found. The ratios of staff dose to patient DAP were found to be within the range of 0.14-3.75 for each procedure. In the second method, cardiologist doses were calculated and compared with the measured values. Scatter doses were measured at the positions of cardiologists from Rando phantom exposures using similar conditions with patient procedures for this purpose. The parameters obtained from these exposures and patient examinations were used to calculate the doses to cardiologists.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization I: Pre-Procedure Overview
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...