Related Experiment Video
Updated: Jun 20, 2026

Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
First experience with the Impella Recover(R) LP 2.5 micro axial pump in patients with cardiogenic shock or undergoing
1Department of Cardiology, UH Leuven, Belgium.
Aim:
To study the feasibility, safety and efficacy of the Recover(R) LP 2.5 assist device in patients scheduled for high risk off-pump coronary bypass surgery, percutaneous coronary intervention or patients in cardiogenic shock.
Methods And Results:
40 patients presenting with cardiogenic shock (n=13) or scheduled for a high risk revascularisation (n=27) were included.36 were selected for safety and feasibility analysis. In 3 patients the pump could not be placed in an adequate position. 5 patients had access related complications. In 9 patients free Hb rose above 80 mg/dl. 3 malfunctions and early device-removal occurred. After device modifications these problems did not recur. CO in the shock group increased significantly: 4.4 l/min+/-1.9 to 4.8 l/min+/-1.2 (p=0.0178).The left ventricular filling pressures decreased in both groups (22 mmHg+/-7.5 to 16 mmHg+/-6 in the shock group, [p=0.0008] and over 6 hours from 14.3 mmHg+/-5.8 to 10 mmHg+/-2.9 in the high-risk revascularisation group,[p=0.0327]).
Conclusions:
The Recover(R) LP 2.5 micro axial pump allows, via percutaneous approach, partial unloading of the left ventricle. The technique is, after design modifications, feasible and safe and results in haemodynamic improvement.