High-Accuracy Correction of 3D Chromatic Shifts in the Age of Super-Resolution Biological Imaging Using Chromagnon
Published on: June 16, 2020
Exploration of chromatic aberration for multiplanar imaging: proof of concept with implications for fast, efficient
Martin Weinigel1, Albert L Kellner, Jeffrey H Price
1Signal Transduction Program, Cancer Center, Burnham Institute for Medical Research, 10901 North Torrey Pines Road, La Jolla, CA 92037, USA.
Abstract:
Image-based autofocus determines focus directly from the specimen (as opposed to reflective surface positioning with an offset), but sequential acquisition of a stack of images to measure resolution/sharpness and find best focus is slower than reflective positioning. Simultaneous imaging of multiple focal planes, which is also useful for 3D imaging of live cells, is faster but requires complicated optics. With color CCD cameras and white light sources commonly available, we asked if axial chromatic aberration can be utilized to acquire multiple focal planes simultaneously, and if it can be controlled through a range sufficient for practical use. For proof of concept, we theoretically and experimentally explored the focal differences between three narrow wavelength bands on a 3-chip color CCD camera with and without glass inserts of various thicknesses and dispersions. Ray tracing yielded changes in foci of 0.65-0.9 microm upon insertion of 12.5-mm thick glass samples for green (G, 522 nm) vs. blue (B, 462 nm) and green vs. red (G-R, 604 nm). On a microscope: (1) With no glass inserts, the differences in foci were 2.15 microm (G-B) and 0.43 microm (G-R); (2) With glass inserts, the maximum change in foci for G vs. B was 0.44 microm and for G vs. R was 0.26 microm; and (3) An 11.3 mm thick N-BK7 glass insert shifted the foci 0.9 microm (R), 0.6 microm (G), and 0.35 microm (B), such that the B and R foci were farther apart (2.1 microm vs. 1.7 microm) and the R and G foci were closer together (0.25 microm vs. 0.45 microm). The slopes of the differences in foci were dependent on thickness, index of refraction, and dispersion. The measured differences in foci are comparable to the axial steps of 0.1-0.24 microm commonly used for autofocus, and focal plane separation can be altered by inserting optical elements of various dispersions and thicknesses. By enabling acquisition of multiple, axially offset images simultaneously, chromatic aberration, normally an imaging pariah, creates a possible mechanism for efficient multiplanar imaging of multiple spectral bands from white light illumination.
More Related Videos
Related Concept Videos
Confocal Fluorescence Microscopy
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...