Impact of skin tone on the performance of a transcutaneous jaundice meter
Stephen Wainer1, Yacov Rabi, Seema M Parmar
1Division of Community Paediatrics, Faculty of Medicine, University of Calgary, Calgary, AB, Canada. wainer@ucalgary.ca
Aim:
To evaluate the performance of the Konica Minolta/Air-Shields JM-103 jaundice meter on the basis of infant skin tone during the early neonatal period.
Methods:
Infants were prospectively categorized into light, medium and dark skin tone groups relative to two reference colours. Transcutaneous bilirubin readings were taken at predetermined intervals through the early neonatal period on a convenience sample of 938 healthy infants > or =37 weeks gestation. Serum bilirubin measurements were drawn routinely with metabolic studies and repeated in the presence of an elevated transcutaneous reading or clinically significant jaundice.
Results:
Multivariate linear regression analysis showed a significant impact on serum and transcutaneous bilirubin agreement by skin tone. Highest precision and lowest bias were observed for medium skin toned infants. Greater disagreement between serum and transcutaneous measurements was noted at serum bilirubin concentrations >200 micromol/L. Insufficient numbers of dark skin toned infants were enrolled to evaluate fully the performance of the jaundice meter for this group.
Conclusion:
The JM-103 jaundice meter displayed good correlation with serum bilirubin concentrations in light and medium skin tone infants, although it showed a tendency to under-read in the lighter skin tone group and to over-read in the darker skin tone group. The device shows excellent performance characteristics for use as a screening device.
Related Concept Videos
Changes in Skin Color: Clinical Perspectives
Albinism
Albinism is a genetic disorder that affects (completely or partially) the coloring of skin, hair, and eyes. The defect is primarily...
Pigmentation
Melanin occurs in two primary forms: eumelanin that provides black and brown pigment and pheomelanin that provides red color. Dark-skinned individuals produce more melanin than those with pale...
Jaundice
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...