Related Experiment Video
Updated: Jun 20, 2026

Translational Rabbit Model of Chronic Cardiac Pacing
Published on: January 6, 2023
Initial experience of pacing with a lumenless lead system in patients with congenital heart disease
Santabhanu Chakrabarti1, Gareth J Morgan, Damien Kenny
1Bristol Congenital Cardiac Centre, Bristol Royal Hospital for Children, Bristol, UK.
Insights
The Medtronic Model 3830 lead offers a safe and effective solution for long-term pacing in congenital heart disease patients, demonstrating acceptable complication and lead displacement rates.
Area of Science:
- Cardiology
- Biomedical Engineering
- Pediatric Electrophysiology
Background:
- Long-term pacing is essential for managing congenital heart disease (CHD).
- Preserving ventricular function and preventing venous occlusion are critical considerations in CHD patients requiring pacing.
- The Model 3830 (SelectSecure) lead offers site-selective pacing with a smaller diameter, designed to address these specific challenges.
Purpose of the Study:
- To evaluate the initial clinical experience with the Model 3830 lead in patients with congenital heart disease.
- To assess the safety and efficacy of site-selective pacing using the Model 3830 lead in this complex patient population.
Main Methods:
- Retrospective analysis of patients who received site-selective Model 3830 lead implantation.
- Data collected from two congenital heart centers between October 2004 and February 2008.
- Included analysis of lead performance, procedural outcomes, and complications.
Main Results:
- 139 leads were implanted in 90 patients (76% children) with various CHD indications.
- Successful pacing was achieved in all procedures, with no procedural mortality.
- Low rates of procedural complications (2) and lead displacement (2.9%) were observed during a median follow-up of 21.8 months.
Conclusions:
- The Model 3830 lead is a safe and effective pacing option for patients with congenital heart disease.
- Despite the technical challenges associated with pacing in CHD, this lead demonstrates acceptable procedural complication and lead displacement rates.
Background:
Long-term pacing is frequently necessary in patients with congenital heart disease (CHD). Preservation of ventricular function and avoidance of venous occlusion is important in these patients. Site-selective pacing with a smaller diameter lead is achievable with the model 3830 lead (SelectSecure, Medtronic Inc., Minneapolis, MN, USA), which was specifically designed to target these complications. We describe our initial experience with the Model 3830 lead in patients with CHD.
Methods:
Retrospective analysis of all patients undergoing site-selective implantation of a Model 3830 lead(s) from two congenital heart centers (Bristol, UK, and Dublin, Ireland) from October 2004 until February 2008.
Results:
We implanted 139 SelectSecure leads (atrial n=70; ventricular n=69) in 90 patients (57 male) with CHD. Median age at implantation: 13.4 years (1.1-59.2 years), median weight: 43 kg. Sixty-nine patients (76%) were children (<18 years). Indications for lead implantation included atrioventricular block (n=55), sinus node disease (n=18), implantable cardiac defibrillator (n=12), antitachycardia pacing (n=4), and cardiac resynchronization (n=1). Twenty-two patients underwent pre-existing lead extraction during the same procedure. All the attempted procedures resulted in successful pacing. One patient had a significantly raised threshold at implantation. There was no procedural mortality. There were two procedural complications. Three patients required lead repositioning for increasing thresholds early postprocedure (<6 weeks). Four leads (2.9%) had displaced on median follow-up of 21.8 months (0.5-42 months).
Conclusions:
The Model 3830 lead is safe and effective in patients with CHD. This is a technically challenging patient group yet procedural complication and lead displacement rates are acceptable.
More Related Videos
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization II: Right Heart Catheterization
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...