Summary
Wideband (WB) acoustic tests, including acoustic reflex thresholds (ARTs) and acoustic transfer functions (ATFs), are more accurate than 1-kHz tympanometry for newborn hearing screening (NHS) referrals. These tests improve middle-ear status assessment for conductive dysfunction and auditory neuropathy.
Area of Science:
- Audiology
- Otoacoustic Emissions
- Hearing Science
Background:
- Newborn hearing screening (NHS) protocols often rely on otoacoustic emissions.
- Accurate assessment of middle-ear status is crucial for identifying hearing impairments in newborns.
- Limitations exist in current screening methods for detecting conductive dysfunction and auditory neuropathy.
Purpose of the Study:
- To evaluate the accuracy of a wideband (WB) aural acoustical test battery in classifying newborn hearing screening outcomes.
- To compare the efficacy of WB acoustic tests (ARTs and ATFs) against 1-kHz tympanometry for NHS referrals.
- To explore the potential of middle-ear assessment to enhance NHS programs.
Main Methods:
- A WB acoustic test battery, including acoustic-reflex thresholds (ARTs) and acoustic-transfer functions (ATFs), was employed.
- Ipsilateral ARTs were measured using broadband noise or tonal activators and clicks, assessing reflex shifts.
- ARTs were analyzed at low (0.8-2.8 kHz) and high (2.8-8 kHz) frequencies using maximum likelihood estimation.
Main Results:
- The median low-frequency ART was significantly elevated (24 dB) in NHS refers compared to passes.
- An optimal combination of ATF and ART tests demonstrated superior prediction of NHS outcomes compared to individual tests.
- WB acoustic tests outperformed 1-kHz tympanometry in classifying NHS outcomes.
Conclusions:
- WB acoustic test batteries, incorporating ARTs and ATFs, offer enhanced accuracy for classifying newborn hearing screening results.
- These WB tests provide a more effective method for identifying middle-ear dysfunction and potential auditory neuropathy in newborns.
- The findings support the integration of comprehensive WB acoustic assessments into newborn hearing screening protocols.