Real-time X-ray Imaging of Lung Fluid Volumes in Neonatal Mouse Lung
Published on: July 18, 2016
Comparison of lung volume measurements by multiple-breath heptafluoropropane washout and computed tomography in small
Hans Proquitté1, Thomas Elgeti, Charles Christoph Roehr
1Clinic of Neonatology (Charité Campus Mitte), Charité Universitätsmedizin, Berlin, Germany. hans.proquitte@charite.de
Background:
Knowledge of lung volume is essential for monitoring and optimizing mechanical ventilation. The aim of this study was to compare lung volume measurements by multiple-breath heptafluoropropane (HFP) washout (MBW) and by respiratory gated computed tomography (CT) in ventilated newborn piglets.
Material/Methods:
In 6 ventilated piglets (age: <12 h, median weight: 945 g) blood gases, respiratory mechanics, and lung volumes were measured in both the supine and prone positions. The measurements were performed in random order. Functional residual capacity (FRC) was measured simultaneously by HFP MBW (FRC(HFP)) using a new infrared mainstream sensor and by CT (FRC(CT)) at the end of inspiration and expiration (multi-slice Toshiba Aquilon 16, Otawara, Japan). Tidal volume (V(T)) was measured both by the Dräger Babylog 8000 ventilator (V(T BL)) and the volume difference of the CT scans (V(T CT)).
Results:
FRC(HFP) (25.2+/-8.5 ml) and FRC(CT) (24.9+/-7.6 ml) correlated strongly (r=0.97) without significant bias. Bland-Altman limits of agreement showed differences between the two methods that ranged from -19.7 to +19.5%. A similar strong correlation without statistically significant bias was found between V(T BL) (8.5+/-2.0 ml) and V(T CT) (9.0+/-2.4 ml) with r=0.91. The limits of agreement were -24.4 and +14.0%. Body position (prone vs. supine) had no significant effect on blood gases, respiratory mechanics, or lung volumes.
Conclusions:
Lung volumes measured in small ventilated lungs by HFP washout and CT are highly correlated and independent of body position. However, the relatively large limits of agreement indicate differences in the two techniques.