Inaccuracy of the Hawksley random zero sphygmomanometer.

E O'Brien1, F Mee, N Atkins

  • 1Blood Pressure Unit, Beaumont Hospital, Dublin, Republic of Ireland.

Lancet (London, England)
|December 15, 1990
PubMed
Summary

The Hawksley random zero sphygmomanometer inaccurately measures blood pressure, consistently underestimating both systolic and diastolic readings. Its current design is unsuitable for hypertension research.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Erratum: Centrality-Dependent Modification of Jet-Production Rates in Deuteron-Gold Collisions at sqrt[s_{NN}]=200  GeV [Phys. Rev. Lett. 116, 122301 (2016)].

Physical review letters·2025
Same author

Quantifying the influence of climate change on pesticide risks in drinking water.

The Science of the total environment·2025
Same author

International consensus statement on the use of ultrasound in hidradenitis suppurativa.

Journal of the European Academy of Dermatology and Venereology : JEADV·2025
Same author

Disentangling Centrality Bias and Final-State Effects in the Production of High-p_{T} Neutral Pions Using Direct Photon in d+Au Collisions at sqrt[s_{NN}]=200  GeV.

Physical review letters·2025
Same author

Review: Sperm cryopreservation in wild small ruminants: morphometric, endocrine and molecular basis of cryoresistance.

Animal : an international journal of animal bioscience·2023
Same author

Measurement of Direct-Photon Cross Section and Double-Helicity Asymmetry at sqrt[s]=510  GeV in p[over →]+p[over →] Collisions.

Physical review letters·2023

Area of Science:

  • Cardiovascular research
  • Medical device accuracy

Background:

  • Accurate blood pressure measurement is crucial for epidemiological and intervention studies.
  • Sphygmomanometers are standard devices for blood pressure monitoring.

Purpose of the Study:

  • To evaluate the accuracy of the Hawksley random zero sphygmomanometer compared to mercury sphygmomanometers.
  • To determine the suitability of the Hawksley device for hypertension research.

Main Methods:

  • Two studies compared readings from the Hawksley device (UK and US models) with mercury sphygmomanometers.
  • Independent observers and device-swapping protocols were used to ensure robust comparison.

Main Results:

  • The Hawksley device showed significant underestimation of both systolic and diastolic blood pressure.

Related Experiment Videos

  • The UK model underestimated systolic by 3.8 mm Hg and diastolic by 7.5 mm Hg.
  • The US model underestimated systolic by 2.6 mm Hg and diastolic by 6.2 mm Hg.
  • Conclusions:

    • The Hawksley random zero sphygmomanometer exhibits significant inaccuracies in blood pressure measurement.
    • Quantitative blood pressure data from studies using this device require re-evaluation.
    • The current design of the Hawksley sphygmomanometer is not recommended for hypertension research.