Related Experiment Video
Updated: Jun 19, 2026

Synchrotron X-ray Microdiffraction and Fluorescence Imaging of Mineral and Rock Samples
Published on: June 19, 2018
Ray tracing flux calculation for the small and wide angle x-ray scattering diffraction station at the SESAME
Wa'el Salah1, M Sanchez del Rio, H Hoorani
1Synchrotron-light for Experimental Science and Application in the Middle East (SESAME), P.O. Box 7, Allan 19252, Jordan. wael.salah@sesame.org.jo
Abstract:
The calculation for the optics of the synchrotron radiation small and wide angle x-ray scattering beamline, currently under construction at SESAME is described. This beamline is based on a cylindrically bent germanium (111) single crystal with an asymmetric cut of 10.5 degrees , followed by a 1.2 m long rhodium coated plane mirror bent into a cylindrical form. The focusing properties of bent asymmetrically cut crystals have not yet been studied in depth. The present paper is devoted to study of a particular application of a bent asymmetrically cut crystal using ray tracing simulations with the SHADOW code. These simulations show that photon fluxes of order of 1.09x10(11) photons/s will be available at the experimental focus at 8.79 keV. The focused beam dimensions will be 2.2 mm horizontal full width at half maximum (FWHM) by 0.12 mm vertical (FWHM).