[Quality control of computed tomography scanners from the viewpoint of radiotherapy planning in Hungary]
Csilla Pesznyák1, Csaba Weisz, Réka Király
1Fovárosi Onkológiai Központ, Budapest, Uzsoki. csilla.pesznyak@freemail.hu
Abstract:
The aim of this publication is to review the requirements necessary for using computed tomography (CT) for radiotherapy treatment planning. The equipments were tested with different CT phantoms. The authors made several measurements for checking the CT number, the quality and the mechanical parameters of CT tables. The CT numbers measured on the equipments of different manufacturers were in quite good agreement with the IAEA requirements. The geometric distortions of CT images are negligible, while the mechanical parameters of CT tables show considerable variety. A quality assurance - quality control protocol is recommended to implement in a safe workflow.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET