Determination of digoxin and digitoxin in whole blood
Elisabeth Leere Oiestad1, Unni Johansen, Mimi Stokke Opdal
1Norwegian Institute of Public Health, Division of Forensic Toxicology and Drug Abuse, P.O. Box 4404 Nydalen, NO-0403 Oslo, Norway. elisabeth.oiestad@fhi.no
Abstract:
A liquid chromatography-tandem mass spectrometry (LC-MS-MS) method has been developed and validated for the determination of digoxin and digitoxin in whole blood samples in autopsy cases. Samples were prepared by liquid-liquid extraction (LLE) with ethyl acetate/heptane/dichloromethane (3:1:1). LC separation was achieved using an Atlantis dC(18)-column (2.1 x 50 mm, 3 microm). The time between injections was 11 min. Mass detection was performed by positive ion mode electrospray LC-MS-MS on the ammonium adducts with two transitions for each analyte and one for the internal standard (digoxin-d(3)). Within-day precision was between 8.3 and 10.8%, between-day precision was between 8.7 and 14.2% and accuracy (bias) was between -17.3 and 11.5%. LOQ was 0.1 nmol/L (0.08 ng/mL), with an accuracy and precision of 19% and -17% (digoxin) and 18% and 3% (digitoxin). Matrix effects ranged from 104 to 117%. Good qualitative correlation with previous findings was achieved for 38 autopsy cases. The median (range, number of cases) B-digitoxin and B-digoxin found with the LC-MS-MS method in this very limited material were 9.3 (3.4-23.8, n = 24) and 5.6 (3.4-26.5, n = 4) nmol/L, respectively.
Related Concept Videos
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Therapeutic Drug Monitoring: Drug Analysis Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Estimation of k and VD of Aminoglycosides
Drug Concentrations: Measurements
Plasma —...
Heart Failure Drugs: Inotropic Agents