Factors influencing intra-articular fluid temperature profiles with radiofrequency ablation
Bojan B Zoric1, Nils Horn, Sepp Braun
1Stetson Powell Orthopedics and Sports Medicine, 191 South Buena Vista Street, Suite 470, Burbank, CA 91505, USA.
Background:
Radiofrequency ablation devices are being used increasingly in arthroscopic surgery. However, there are concerns that excessive temperatures may damage the articular cartilage. The purpose of this study was to investigate the temperature profiles that occur within the glenohumeral space with the use of one commercially available radiofrequency ablation probe.
Methods:
Ten fresh-frozen human cadaver shoulder specimens were used. Intra-articular temperatures were measured at different time intervals over a two-minute period at a distance of 1, 3, 5, and 10 mm away from the probe. The radiofrequency probe was activated throughout the range of machine power settings, and irrigation fluid flow was varied (no flow, a flow at 60 mm Hg without suction, and a flow at 60 mm Hg with suction).
Results:
Temperatures deleterious to articular cartilage chondrocytes (i.e., those in excess of 50 degrees C) were seen with an increased duration of application, a decreased distance between the thermometer and the probe, and a decreased irrigation fluid flow rate. The highest recorded irrigation fluid temperature reached >80 degrees C after two minutes in a no-flow setting. The flow rate was found to be the most significant predictor of intra-articular temperature profiles. The various machine power settings had no apparent influence on temperature, meaning that higher probe settings are not necessarily associated with higher temperature profiles.
Conclusions And Clinical Relevance:
These results demonstrate the importance of the management of the irrigation fluid flow rate across the joint during arthroscopic procedures that involve radiofrequency ablation. Even short intervals of limited flow could lead to supraphysiological temperature profiles and potentially to cartilage damage.
More Related Videos
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Related Concept Videos
Factors Affecting Body Temperature
Factors may include:
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...