Related Experiment Video
Updated: Jun 19, 2026

Rapid Mix Preparation of Bioinspired Nanoscale Hydroxyapatite for Biomedical Applications
Published on: February 23, 2017
High revision rate of hydroxyapatite-coated ABG-I prosthesis
Young-Yool Chung1, Sung-Chan Ki, Kwang-Young So
1Department of Orthopedic Surgery, Gwangju Christian Hospital, Gwangju, Republic of Korea.
Background:
The aim of this study was to investigate the causes of the high revision rate of the hydroxyapatite-coated Anatomique Benoist Girard (ABG-I) prosthesis.
Methods:
We performed 204 total hip arthroplasties (THAs) at our hospital between March 1992 and December 1996. Of the 204 THA patients, 129 were followed up; the mean duration of follow-up was 12.2 years (range 10.0-14.5 years). There were 113 male patients and 16 female patients, with an average age of 53 years (30-83 years) at the time of surgery. The reasons for THA were avascular necrosis in 119 hips, acetabular dysplasia in 8 hips, and traumatic arthritis in 2 hips.
Results:
The Harris hip score was 47.3 preoperatively and 86.4 at the last follow-up. The linear polyethylene wear was an average of 0.29 mm/year. Acetabular osteolysis was seen in 113 cases (88%). Altogether, 61 (47.2%) acetabular cups were revised for aseptic loosening in 38 hips, polyethylene wear and osteolysis in 20 hips, recurrent dislocation in 2 hips, and deep infection in 1 hip. Femoral osteolysis was observed in 100 cases (77.5%). The femoral stem was revised in 4 hips (3.1%).
Conclusion:
We observed that the fatal detriment to ABG-I acetabular cup survival in long-term follow-up was periacetabular osteolysis followed by aseptic loosening.
More Related Videos
07:14Synthesis of Graphene-Hydroxyapatite Nanocomposites for Potential Use in Bone Tissue Engineering
Published on: July 27, 2022
09:35Distinctive Capillary Action by Micro-channels in Bone-like Templates can Enhance Recruitment of Cells for Restoration of Large Bony Defect
Published on: September 11, 2015