Dosimetric comparison of ring and ovoid applicators
N Tuncell1, A Toy, A N Demiral
1Department of Radiation Oncology, Akdeniz University Medical School, Antalya, Turkey.
Ring applicators significantly reduce rectal dose compared to ovoid applicators in vaginal cuff brachytherapy. However, this comes with a notable decrease in the reference volume size for treatment planning.
Area of Science:
- Medical Physics
- Radiation Oncology
- Brachytherapy
Background:
- Postoperative vaginal cuff irradiation is a crucial treatment modality for gynecological cancers.
- Accurate dosimetry is essential to maximize tumor control while minimizing toxicity to surrounding organs at risk, such as the rectum.
Purpose of the Study:
- To compare the dosimetric characteristics of different vaginal applicators (ovoids and rings) used in vaginal cuff brachytherapy.
- To evaluate the impact of applicator size and configuration on dose distribution to the vaginal cuff and rectum.
Main Methods:
- A dosimetric study using the Plato Treatment Planning System (TPS) was conducted.
- Standard ovoid and ring applicators of various sizes and angles were modeled.
- Dose calculations were performed for defined reference points, including the vaginal cuff (V(c)) and multiple rectal reference points (R(1)-R(5)).
Main Results:
- Ovoid applicators resulted in a significantly higher rectal reference dose (R(1)) compared to ring applicators (43.1% vs. 28.6%).
- Increasing applicator size led to an increase in rectal dose and the thickness of the reference isodose.
- The magnitude of the reference volume was independent of ring angle but decreased significantly with ring applicators compared to ovoids.
Conclusions:
- Ring applicators offer a dosimetric advantage by reducing the maximum rectal reference dose compared to ovoid applicators when used with recommended dwell positions.
- The reduction in rectal dose with ring applicators is associated with a significant decrease in the reference treatment volume.
- Careful selection of applicator size and type is necessary to optimize dose delivery and minimize toxicity in vaginal cuff brachytherapy.