Related Experiment Video
Updated: Jun 19, 2026

Preparation of Food Samples Using Homogenization and Microwave-Assisted Wet Acid Digestion for Multi-Element Determination with ICP-MS
Published on: December 22, 2023
[Determination of trace elements in electrical absorption prospecting polyform sample by inductively coupled plasma
Yi-Hua Shi1, Zhong-Ping Yang, Jian-Hui Huang
1Guilin Research Institute of Geology for Mineral Resources, Guilin 541004, China. syhua2000@163.com
Abstract:
An ICP-MS method was established for the determination of sixteen trace elements, V, Cr, Mn, Co, Ni, Cu, Zn, Ga, Nb, Mo, Ag, Cd, Au, Tl, Pb and Bi in electrical absorption prospecting polyform. Three methods for polyform samples (ashing method, extraction by HNO3 + H2O2 and digestion with aqua regia) were compared and the results showed that the second method is the best one. The best operational paramenters of X series ICP-MS were confirmed, the inner standard 103Rh and 185Re were selected for the determination of elements, and analysis of isotopes interference correction equations was established. Satisfactory linearity of working curves of the sixteen trace elements was obtained, giving all their correlation coefficients over 0.999 8. The determination limit of the analytes was in the range of 0.001-2.2 microg x g(-1). The precision was 1.39%-4.84%, and the recoveries were between 94.86% and 105.2%. The method is sensitive, quick and simple and has been applied to the analysis of a great number of polyform samples.
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...