Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Adjustable shunt valve-induced magnetic resonance imaging artifact: a comparative study
Ahmed K Toma1, Andrew Tarnaris, Joan P Grieve
1Victor Horsley Department of Neurosurgery, National Hospital for Neurology and Neurosurgery, Queen Square, London, United Kingdom. ahmedktoma@yahoo.com
Adjustable shunt valves cause magnetic resonance imaging (MRI) artifacts. The Strata II valve produced larger artifacts than the proGAV valve, especially on diffusion-weighted (DW) MRI sequences, potentially obscuring brain conditions.
Area of Science:
- Neurosurgery
- Radiology
- Medical Device Engineering
Background:
- Adjustable shunt valves are implanted to manage cerebrospinal fluid (CSF) pressure.
- These valves can cause artifacts in Magnetic Resonance Imaging (MRI), potentially affecting diagnostic accuracy.
- Understanding and quantifying these artifacts is crucial for patient management.
Purpose of the Study:
- To compare magnetic resonance imaging (MRI) artifacts induced by two types of implanted adjustable shunt valves: proGAV and Strata II.
- To evaluate artifact prevalence across different MRI pulse sequences, including spin echo (SE), diffusion-weighted (DW), and gradient echo (GRE).
Main Methods:
- MR images from 8 patients with proGAV and 6 patients with Strata II valves were analyzed.
- Artifact areas and volumes were quantified for different pulse sequences and imaging planes.
- Statistical comparisons were made between the two valve types.
Main Results:
- The Strata II valve induced significantly larger artifacts than the proGAV valve in both spin echo (SE) and diffusion-weighted (DW) MRI sequences.
- Artifacts were more pronounced on DW images compared to SE sequences for both valve types.
- Specific quantitative differences were noted: Strata II vs. proGAV in SE (29,761 vs. 2450 mm(3), p=0.003) and DW (100,138 vs. 38,955 mm(3), p=0.025).
Conclusions:
- Adjustable shunt valves, particularly the Strata II, can create significant MRI artifacts that may obscure underlying brain pathology.
- The choice of shunt valve type and implantation site should consider potential MRI artifact generation.
- Further development of less artifact-inducing adjustable shunt valves is warranted, especially for applications involving diffusion-weighted imaging.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...