Are dialysis devices usable as ozone gas exchangers?
Valter Travagli1, Iacopo Zanardi, Alessandro Gabbrielli
1Dipartimento Farmaco Chimico Tecnologico, Università degli Studi di Siena, Siena, Italy.
Abstract:
A study aimed to compare the efficiency of the ozone transfer of four hydrophilic dialysis filters, and one hydrophobic gas-exchange device (GED) has been performed. Obviously, the former should be specifically used only for dialysis. Unfortunately, some clinicians incautiously use them as GEDs. It has been shown that: (i) dialysis filters present a wide range of gas-exchange yield (from 0 up to 70%), often related to variability according to the treatment time; (ii) by scanning microscopy, it has been noticed that hollow fibers are somewhat altered by ozone; and (iii) because their constitutive materials may not be ozone-resistant, they may release toxic compounds harmful for the patients. On the contrary, the appropriate GED is ozone-transfer efficient, is ozone-resistant, and is suitable for blood autotransfusion and ozonation.
Related Concept Videos
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Hemodialysis I: Introduction
Hemodialysis II: Procedure and Complications