Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
Published on: May 3, 2018
Blood pressure measurement: arm circumference and cuff size availability
Eugenia Velludo Veiga1, Edna Aparecida Moura Arcuri, Lyne Cloutier
1Escola de Enfermagem de Ribeirão Preto, Universidade de São Paulo, Brazil. evveiga@eerp.usp.br
Insights
Accurate blood pressure (BP) measurement requires correct cuff size. Most inpatients have arm circumferences that do not fit standard BP cuffs, leading to inaccurate readings.
Area of Science:
- Clinical Medicine
- Biomedical Engineering
- Cardiovascular Research
Background:
- Accurate blood pressure (BP) measurement is crucial for patient care.
- The American Heart Association (AHA) provides guidelines for appropriate BP cuff size selection based on arm circumference (AC).
- Standard BP cuffs may not accommodate the wide range of patient arm sizes, potentially compromising measurement accuracy.
Purpose of the Study:
- To determine the arm circumference (AC) of inpatients.
- To calculate the appropriate BP cuff width (CW) for these ACs based on AHA recommendations (CW/AC ratio of 0.40).
- To assess the availability of suitable BP cuff sizes in clinical settings.
Main Methods:
- Measured the AC of 81 inpatients' right arms.
- Calculated the required CW for each AC using the AHA 0.40 ratio.
- Compared calculated CWs to the standard 12 cm wide cuff available in clinics.
Main Results:
- Patient ACs ranged from 17.5 to 40.5 cm, requiring CWs from 6 to 16 cm.
- The standard 12x23 cm cuff fit only 17.3% of patients (AC 32.5-34.3 cm).
- The standard 12 cm wide cuff was unsuitable for 82.7% of patients, risking inaccurate BP readings.
Conclusions:
- A significant discrepancy exists between available standard BP cuff sizes and the measured ACs of inpatients.
- The limited availability of diverse BP cuff sizes poses a challenge, leading to frequent use of ill-fitting cuffs.
- Inappropriate cuff sizes can result in overestimation or underestimation of blood pressure, impacting clinical decisions.
Abstract:
To avoid inaccurate blood pressure (BP) readings, the American Heart Association (AHA) recommends cuff width (CW) encircling 40% of the arm circumference (AC) and cuff length at least 80-100%. This study aimed to identify inpatients AC, the corresponding cuff size and the cuff size availability. In total, 81 AC were measured in the right arm. The cuff sizes to fit them were calculated according to AHA AC/CW width 0.40 ratio. The AC varied from 17.5 to 40.5 cm and the corresponding cuff width from 6 to 16 cm. The standard cuff 12 by 23 cm, the only size available in the clinics, was appropriate for only 17.3% of the subjects, whose AC varied between 32.5 and 34.3 cm. The lack of availability of different cuff sizes continues being a challenging problem to be faced. The standard cuff available, 12 cm large, did not fit 82.7% of the identified AC, resulting in over or underestimated BP registers.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)