Impact of occluder device type on success of percutaneous closure of atrial septal defects--a medium-term follow-up

Michael Becker1, Dorothee Frings, Jörg Schröder

  • 1University Hospital RWTH Aachen, Department of Cardiology, Aachen, Germany. mibecker@ukaachen.de

Insights

The Amplatzer occluder demonstrated superior performance in percutaneous closure of patent foramen ovale and atrial septal defects, showing fewer residual shunts and lower thromboembolic event rates compared to Starflex and Cardiastar devices.

Area of Science:

  • Cardiovascular Medicine
  • Interventional Cardiology
  • Medical Device Technology

Background:

  • Patent foramen ovale (PFO) and atrial septal defects (ASDs) are common cardiac anomalies.
  • Percutaneous closure is a standard treatment for selected patients, particularly those with paradoxical embolism.
  • Occluder device choice may influence procedural outcomes and long-term efficacy.

Purpose of the Study:

  • To compare the efficacy and safety of different occluder devices (Amplatzer, Starflex, Cardiastar) for PFO and ASD closure.
  • To evaluate rates of residual shunts and thromboembolic events at medium-term follow-up.
  • To identify risk factors for adverse events after percutaneous closure.

Main Methods:

  • A retrospective study of 302 patients undergoing percutaneous PFO or ASD closure between 1997 and 2006.
  • Patients received either Amplatzer, Starflex, or Cardiastar occluders.
  • Follow-up included transesophageal echocardiography at 1, 3, and 6 months, and assessment of thromboembolic events over a mean of 31 months.

Main Results:

  • The Amplatzer group had significantly fewer residual shunts at 6 months compared to Starflex and Cardiastar (P=0.0005).
  • Occluder type was a significant risk factor for residual shunts (P=0.0033).
  • The annual recurrence rate of thromboembolic events was lower with Amplatzer (1.3%) versus Starflex (3.9%) and Cardiastar (3.6%) (P=0.0467). Atrial septal aneurysm was a significant risk factor for thromboembolic events (P=0.0168).

Conclusions:

  • Percutaneous closure of PFO/ASD is a safe procedure with low complication rates.
  • Significant differences exist among Amplatzer, Cardiastar, and Starflex occluders regarding defect closure and thromboembolic event recurrence.
  • Amplatzer devices appear to offer better long-term outcomes in terms of shunt closure and thromboembolic event prevention.