Ultrasound-based navigation and 3D CT compared in acetabular cup position
Related Experiment Video
Updated: Jun 19, 2026

In Vivo Quantification of Hip Arthrokinematics during Dynamic Weight-bearing Activities using Dual Fluoroscopy
Published on: July 2, 2021
Olaf Hasart1, Berry M Poepplau, Patrick Asbach
1Charité Orthopaedic Department, Center for Musculoskeletal Surgery, Charité Universitätsmedizin, 10117 Berlin, Germany.
Abstract:
Intraoperative landmarks are used in image-free navigation systems. The ultrasound-based navigation systems try to overcome the problems of positional deviation associated with soft tissue. Our study analyzed the accuracy of ultrasound-based navigation of cup positioning compared with postoperative 3-dimensional (3D) computed tomography scans of cup positioning. Twenty-five ultrasound-navigated total hip arthroplasties (THAs) were analyzed. The difference between the intraoperative cup orientation (navigation) and the postoperative cup position (CT) was evaluated. The average difference between intraoperative navigation and postoperative CT measurements was 2.8 degrees (SD+/-1.8 degrees ) for abduction and 2.2 degrees (SD+/-1.6 degrees ) for anteversion. Therefore, we recommend ultrasound-based navigation as an exact tool for cup positioning in THA.