Related Experiment Video
Updated: Jun 19, 2026

Synthesis and Operation of Fluorescent-core Microcavities for Refractometric Sensing
Published on: March 13, 2013
Six orders of magnitude dynamic range in capillary electrophoresis with ultrasensitive laser-induced fluorescence
Colin D Whitmore1, David Essaka, Norman J Dovichi
1Department of Chemistry, University of Washington, Seattle, WA 98195-1700, USA. cwhitmor@u.washington.edu
Abstract:
An ultrasensitive laser-induced fluorescence detector was used with capillary electrophoresis for the study of 5-carboxy-tetramethylrhodamine. The raw signal from the detector provided roughly three orders of magnitude dynamic range. The signal saturated at high analyte concentrations due to the dead time associated with the single-photon counting avalanche photodiode employed in the detector. The signal can be corrected for the detector dead time, providing an additional order of magnitude dynamic range. To further increase dynamic range, two fiber-optic beam-splitters were cascaded to generate a primary signal and two attenuated signals, each monitored by a single-photon counting avalanche photodiode. The combined signals from the three photodiodes are reasonably linear from the concentration detection limit of 3 pM to 10 microM, the maximum concentration investigated, a range of 3,000,000. Mass detection limits were 150 yoctomoles injected onto the capillary.
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Electrophoresis: Overview
There...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...