High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Supported liquid extraction in combination with LC-MS/MS for high-throughput quantitative analysis of hydrocortisone
Shari Wu1, Wenkui Li, Tapan Mujamdar
1Department of Drug Metabolism and Pharmacokinetics, Novartis Institutes for Biomedical Research, One Health Plaza, East Hanover, NJ 07936, USA.
Abstract:
A high-throughput LC-MS/MS bioanalytical method was developed and validated for the determination of hydrocortisone in mouse serum via supported liquid extraction (SLE) in a 96-well plate format. Although sample extracts from SLE result in similar matrix effects compared with conventional liquid-liquid extraction (LLE), greater analyte extraction recovery and much higher analysis throughput for the quantitative analysis of hydrocortisone in mouse serum were obtained. The current LC-MS/MS method was validated for a concentration range of 2.00-2000 ng/mL for hydrocortisone using a 0.100 mL volume of mouse serum. The intra- and inter-day precision and accuracy of the quality control samples at low, medium and high concentration levels showed < or =12.9% CV and -3.4-6.2% bias for the analyte in mouse serum.
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process