Related Experiment Video
Updated: Jun 19, 2026

Pediatric Animal Model of Extracorporeal Cardiopulmonary Resuscitation After Prolonged Circulatory Arrest
Published on: May 26, 2023
Extensive injury after use of a mechanical cardiopulmonary resuscitation device
J Wind1, S C A M Bekkers, L J H van Hooren
1Department of Surgery, Maastricht University Medical Centre, Maastricht, The Netherlands. t.wind@mumc.nl
Abstract:
We report a case of a 49-year-old woman with a ruptured liver and spleen found at autopsy, which may have been related to the use of a mechanical cardiopulmonary resuscitation (CPR) device (AutoPulse, ZOLL Medical Corporation, Chelmsford, Mass). She was admitted because of an out-of-hospital resuscitation, and under the suspicion of a pulmonary embolism, a thrombolytic agent was administered. Despite prolonged continuation of mechanical CPR, she died of persistent asystole. The evidence for improved outcomes after the use of a mechanical CPR device during resuscitation is still scarce. To prevent the unique complications reported here, regular checking of proper position of the chest band during resuscitation is advised.